Search

Meatest MTE Laboratory Impedance Standards

Laboratory impedance standards for resistance and capacitance, 10 MΩ–10 GΩ, 0.005% accuracy, low time constant and traceable calibration.

Contact information

Resistance and Capacitance
  • Four package types for various applications
  • Custom nominal values and terminals upon request
  • Low time constant

MTE series resistance and capacitance standards are designed for calibration of ohmmeters, LCR meters and insulation testers. Alternatively, MTE standards can be used as transfer standards, keeping traceability between primary and secondary laboratories.
MTE standards come in four terminal configurations to fit every application.
OPEN and SHORT standards for zero calibration of LCR meters as well as custom nominals and terminal configurations are available upon request.
 
Specifications below describe 1-year absolute accuracy of this product including long-term stability, linearity and reference standard measurement uncertainty as well as ambient conditions within specified limits.

  • Reference temperature: 21 °C ± 4 °C
  • Operating temperature: 0 °C to +50 °C
  • Dimensions (W x H x D): 125 x 60 x 105 mm
  • Resistance Construction:
    • Up to 1 MΩ: Foil resistor
    • Above 1 MΩ: Ceramic resistor
  • Frequency range: DC - 20 kHz
  • Available packages:
    • Up to 1 MΩ: A, B and Cc
    • Above 1 MΩ: HV; DC applications only
  • Reference voltage: 1 kV for range 10 MΩ - 10 GΩ
  • Voltage coefficient: 0.5 ppm/V (typical) for range 10 MΩ - 10 GΩ