The TA043 is an active differential probe suitable for high common-mode voltage measurement applications up to ±700 V (DC + peak AC).
It is capable of acquiring higher speed signals of up to 100 MHz.
This differential probe extends the functionality of standard single-ended input oscilloscopes to allow a safe and accurate method of making high-voltage differential measurements.
Applications include making safe measurements in power circuit applications and acquisition of low speed balanced differential signals found in serial communications buses.
TA043 differential oscilloscope probe specifications
Bandwidth: DC to 100 MHz (-3 dB)
Attenuation ranges: 10:1 / 100:1
DC accuracy: ±2%
Input impedance: 4 MΩ ∥ 7 pF each side to ground
Input
Differential input voltage range (10:1): ±70 V (DC + peak AC) or 70 V RMS
Differential input voltage range (100:1): ±700 V (DC + peak AC) or 500 V RMS
Common mode voltage range: ±700 V (DC + peak AC) or 500 V RMS(10:1 and 100:1)
Absolute maximum voltage(differential or common-mode): 1000 V RMS CAT III (10:1 and 100:1)
Output
Swing: ±7 V into 50 kΩ loadOffset (typical): <±5 mV
Noise (typical): 0.9 mV RMS
Source impedance (typical): 50 Ω
CMRR (typical): 85 dB @ 50 Hz; 55 dB @ 1 MHz Ambient operating temperature: -10 to +40 °C
Power requirements: 4xAA cells or 6 V DC / 200 mA mains adaptor or regulated 9 V DC / 120 mA mains adaptor
Length of input leads: 30 cm
Length of BNC lead: 90 cm
Weight: 500 g (inc. probes and PVC jacket)
Dimensions: 202 mm x 83 mm x 38 mm