Haefely FP-EFT 100M2, 3-phase Coupling / Decoupling Network (CDN)
Bilder
  • Haefely FP-EFT 100M2, 3-phase Coupling / Decoupling Network (CDN)

Haefely FP-EFT 100M2, 3-phase Coupling / Decoupling Network (CDN)

Art. nr:

Begär offert

 
 

Produktbeskrivning

IEC/EN 61000-4-4, ANSI C62.41 &
C.37.90.1
  • 5/50 ns impulse shape
  • 8 kV Burst max. impulse voltage
  • Max. AC voltage:
690 V AC (phase – phase),
400 V AC (phase – neutral)
  • 100 A per phase EUT current
The FP-EFT 100M2 coupling/decoupling network is fully compliant with the requirements of IEC/EN 61000-4-4, as well as tests according to ANSI/IEEE C62.41 and C.37.90.1 power lines.
 
IEC / EN and ANSI standards cover burst testing of single and three phase AC and DC power ports. They include recommendations for the test equipment, the test set-ups and the test procedures. The IEC 61000- 4-4 Edition 3: 2015 now defines the impulse shape not only at the generator output but also at the CDN output. This Edition 3 requires a pure common mode coupling only. This means that the burst signals are coupled simultaneously into all paths at the same
time. Because of this no coupling path selection during testing is necessary. This speeds up the test time.
 
The FP-EFT 100M2 couples the burst into an EUT while
preventing the impulses from polluting the three phase power supply. The FP-EFT 100M2 has manual coupling path selection to verify the functionality of each path as recommended in the IEC 61000-4-4 Edition 3: 2015.
 
This can also be used to select different coupling paths
according to ANSI standards which requires a path by path burst test. Used together with the AXOS5/AXOS8, the FP-EFT 100M2 can be used to inject EFT/burst impulses on the three phase supply.

 
 

Köp-till

Burst Verification Adapter
Burst Verification Adapter
Verification adapter to check the FP-EFT 100M2